Gauge Pressure Calibrator, 200 kPa, Standard/Metric, 1/4 in. NPT Female Thread

$5,287
Quantity

Description

The Yokogawa CA700-E-01-U2-P2 Gauge Pressure Calibrator, 200 kPa has a basic accuracy of 0.01% of reading on pressure measurement and 0.015% of reading on Current/Voltage source/measurement. It also boasts a resolution and pressure range.

Pressure is one of the most important and commonly measured forces in industrial plants and sites. Other commonly measured forces, such as temperature, flow and level, can all be derived from pressure. Because of this, pressure transmitters are especially crucial for maintaining the health and safety of plant operations. For a pressure transmitter to perform reliably, periodic calibration of these devices is absolutely necessary.

Product Details
YO85804782YC
  • 1 - Gauge Pressure Calibrator, 200 kPa
  • 1 - Set of Lead Wires with Alligator Clips (Red and Black/1.7m), 98064
  • 6 - Alkaline AA Batteries
  • 1 - 1/4 in. NPT Male Thread to 1/8 in. NPT Female Thread Conversion Connector
  • 1 - Accessory Case, B9108XA
  • 1 - Shoulder Strap
  • 1 - User Manual
  • 1 - Startup Guide

Product Features:

  • Supports customizable, bulit-in "AS-FOUND" and "AS LEFT" routines
  • Can be used as a data logger or USB storage
  • Integration with FieldMate 3.03 and pressure modules for all-in-one calibration system
  • Utilizes same high quality Yokogawa sensor technology
  • Different pressure ranges allow maximum resolution
  • Reliable, rugged, and battery-powered
  • IP54 dust and water proof case for harsh enviroments
  • Light-weight and built-in stand
  • Pressure type: Gauge
  • Measurement range:
  • Positive pressure: 0 to 200 kPa
  • Negative pressure: -80 to 0 kPa
  • Measurement display range: To 240.000 kPa
  • Resolution: 0.001 kPa
  • Measurement accuracy (6 months after calibration/Tested after zero calibration):
  • Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.003 kPa); 0 to 20 kPa: ±0.005 kPa
  • Negative pressure: ±(0.2% of reading + 0.080 kPa)
  • Measurement accuracy (1 year after calibration/Tested after zero calibration):
  • Positive pressure 20 to 200 kPa: ±(0.01% of reading + 0.010 kPa); 0 to 20 kPa: ±0.012 kPa
  • Negative pressure: ±(0.2% of reading + 0.090 kPa)
  • Response time: 2.5 s or less
  • Allowable input: 2.7 kPa abs to 500 kPa gauge
6 other products in the same category:

Reference: AE14763789A8

Brand: AEMC

AEMC 8333

PowerPad III Thee-Phase Power Quality Meter

Price $2,499
More
In stock

Reference: FL41605596FI

Brand: Fluke

Fluke II900

Industrial Acoustic Imager

Price $17,901